Paper
A 174.7-dB FoM, 2nd-Order VCO-Based ExG-to-Digital Front-End Using a Multi-Phase Gated-Inverted-Ring Oscillator Quantizer
Reviewer: Xinao Ji
Full Review: Read the Review Document (PDF)
Introduction
This paper presents a direct-digitization front-end for ExG biopotential acquisition. Fabricated in 65 nm CMOS, the chip combines a second-order VCO-based architecture with a multiphase gated-inverted-ring oscillator (GIRO) quantizer to achieve low noise and high linearity over a wide input range, supporting the acquisition of weak biopotential signals in the presence of motion artifacts.
The design reduces the quantizer’s sensitivity to mismatch and uses time-domain signal processing to make part of the circuit’s power consumption vary with input amplitude. An input-impedance boosting circuit reduces the loading imposed on the electrodes.
Key Features
- Combines VCO-based time-domain integration and multiphase quantization to achieve second-order noise shaping
- Uses a GIRO quantizer to reduce mismatch sensitivity and improve linearity
- Supports a 400 mVpp input range, providing headroom for motion artifacts and interference
- Uses input-dependent power consumption to reduce energy use when large artifacts are absent
- Combines AC coupling, chopping, and input-impedance boosting for biopotential acquisition
- Demonstrates ECG, EOG, and EMG measurements, including ECG recording in the presence of motion artifacts
Key Performance
| Parameter | Performance |
|---|---|
| Technology | 65 nm CMOS |
| Core area | 0.075 mm² |
| Analog / digital supply | 1.2 V / 0.8 V |
| Noise-shaping order | Second order |
| Input range | 400 mVpp |
| Sampling frequency | 200 kHz |
| Signal bandwidth | 1 kHz |
| Power consumption | 4.25–5.8 µW |
| Peak SNDR | 92.3 dB |
| Dynamic range | 92.3 dB |
| SFDR | 110.3 dB |
| Input-referred noise density | 110 nV/√Hz |
| Input impedance (DC / bandwidth edge) | 60 / 50 MΩ |
| CMRR | 100.2 dB |
| Schreier FoM | 174.7 dB |
Summary
This work illustrates how time-domain quantizer design, second-order noise shaping, and electrode-interface optimization can work together in a direct-digitization biopotential front-end. The GIRO quantizer addresses mismatch-related linearity limitations, input-dependent power consumption reduces energy use during normal recording, and input-impedance boosting supports the electrode interface.
The wide input range helps prevent front-end saturation in the presence of motion artifacts; it does not directly remove those artifacts. Its benefit is to preserve the recorded waveform, including the underlying weak biopotential signal, for subsequent processing.
For a detailed analysis of the second-order VCO architecture, multiphase GIRO quantizer, input-dependent power consumption, and input-impedance boosting circuit, please refer to the full review document.